| PTB110 Barometer |
| |
|
The PTB110 Barometer is designed both for accurate barometric pressure measurements at room temperature and or general environmental pressure monitoring over a wide temperature range. It uses Vaisala''s BAROCAP(R) silicon capacitive sensor to measure barometric pressure over a 500 to 1100 millibar(hPa) range with an output of 0 to 2.5v and a low current consumption, making it suitable for use with a wide range of data loggers. Features:
> Accuracy of +/-0.3 mb (hPa) @ +20 degC
> Long-term stability +/-0.1 mb (hPa) per year
> Output voltage 0 to 2.5VDC = 500 to 1100 mb (hPa)
> Current consumption less than 4mA
> On/off control with external trigger > Mountable on a 35mm wide DIN rail
> NIST traceable calibration (certificate is included)
The PTB110 is a popular choice for wind-assessment and general-meteorology applications where a barometer is required.
|
Specification Summary:
| Accuracy: | +/-0.3mb(hPa) at +20C |
| | +/-0.6mb(hPa) at 0C to +40C |
| | +/-1.0mb(hPa) at -20C to +45C |
| | +/-1.5mb(hPa) at -40C to +60C |
| Linearity: | +/-0.25mb(hPa) |
| Hysteresis: | +/-0.03mb(hPa) |
| Repeatability: | +/-0.03mb(hPa) |
| Calibration Uncertainty: | +/-0.15mb(hPa) |
| Long-Term Stability: | +/-0.1mb(hPa) per year |
| Supply Voltage Range: | 10 to 30V DC |
| Current Consumption: | < 4mA active, < 1uA quiescent |
| Settling Time: | 1s to reach full accuracy after power-on |
| Response Time: | 500mS to reach full accuracy after a pressure step |
| Operating Temperature Range: | -40C to +60C (humidity: non-condensing) |
| Housing: | ABS/PC with aluminium back-plate |
| Sealing: | IP32 - The PTB110 MUST be installed in a VENTED weather-proof enclosure |
| Dimensions: | 68.4 x 97.8 x 28.1mm (Weight: 90g) |
| CE/EMC: | CE marked (complies with EN61326-1) |
|
|